equipment
Electronic Load Molecule
Public Made by Adomby adom
Siglent SDL1020X-E electronic load equipment prototype: reconstructed carrier STEP/GLB, constant-current dashboard, serial-pinned SCPI driver, control-panel integration guide and purchase budget.
main
Validation record
2026-10-07. Software and CAD checks only; no instrument, fabricated carrier or physical DUT was available.
- Five unit tests pass: disconnected ON rejection; negative/out-of-range/NaN/infinite/bool/non-numeric current rejection; lease expiry OFF; simulation tagging; transport failure clears stale measurements and marks input state unknown.
- Installed package tests also pass. User skill installed to both Claude and Codex skill trees; source is explicit about qualification limits.
- HTTP checks pass: invalid current returns 400, foreign Origin returns 403, non-JSON control returns 415, explicit trusted proxy Origin succeeds, OFF succeeds in simulation.
- Browser verified model render, orbit/explode/reset, 0.1/0.5/1 A requests, ON/OFF, over-limit error feedback, and no console errors after the fixes.
- GLB instrument-only bounding box checked against 256 x 410 x 115 mm within 0.01 mm. 34 named bodies; STEP and GLB regenerated from the same source. This verifies model scale, not dimensional truth of estimated features.
- Public wiki render inspected in browser. Source and installable package are separate published layers.
Not verified: real USBTMC/VXI-11 enumeration, SCPI command acceptance on the received firmware, current accuracy, actual protection thresholds/delays, remote sense, noise/ripple, load transients, current/voltage calibration, USB grounding, support strength, pin retention, robot picking, vent clearance, dock clearance or production layout import. No purchasing action performed.
# Validation record
2026-10-07. Software and CAD checks only; no instrument, fabricated carrier or physical DUT was available.
- Five unit tests pass: disconnected ON rejection; negative/out-of-range/NaN/infinite/bool/non-numeric current rejection; lease expiry OFF; simulation tagging; transport failure clears stale measurements and marks input state unknown.
- Installed package tests also pass. User skill installed to both Claude and Codex skill trees; source is explicit about qualification limits.
- HTTP checks pass: invalid current returns 400, foreign Origin returns 403, non-JSON control returns 415, explicit trusted proxy Origin succeeds, OFF succeeds in simulation.
- Browser verified model render, orbit/explode/reset, 0.1/0.5/1 A requests, ON/OFF, over-limit error feedback, and no console errors after the fixes.
- GLB instrument-only bounding box checked against 256 x 410 x 115 mm within 0.01 mm. 34 named bodies; STEP and GLB regenerated from the same source. This verifies model scale, not dimensional truth of estimated features.
- Public wiki render inspected in browser. Source and installable package are separate published layers.
Not verified: real USBTMC/VXI-11 enumeration, SCPI command acceptance on the received firmware, current accuracy, actual protection thresholds/delays, remote sense, noise/ripple, load transients, current/voltage calibration, USB grounding, support strength, pin retention, robot picking, vent clearance, dock clearance or production layout import. No purchasing action performed.