main
John Lauer
Deliver Board34 R10 prototype review, current simulator twin and 4m34s narrated process cut
47014bd
from pathlib import Path
from reportlab.platypus import SimpleDocTemplate,Paragraph,Spacer,Table,TableStyle,PageBreak
from reportlab.lib.styles import getSampleStyleSheet
from reportlab.lib import colors
r=Path(__file__).resolve().parents[1]
d=r/"docs/project-page"
d.mkdir(exist_ok=True)
styles=getSampleStyleSheet();story=[]
def p(t,style='BodyText'):story.append(Paragraph(t,styles[style]));story.append(Spacer(1,9))
def table(rows):
t=Table([[Paragraph(str(c),styles['BodyText']) for c in row] for row in rows],colWidths=[155,345]);t.setStyle(TableStyle([('VALIGN',(0,0),(-1,-1),'TOP'),('BACKGROUND',(0,0),(-1,0),colors.HexColor('#e6f3f3')),('GRID',(0,0),(-1,-1),.4,colors.grey),('LEFTPADDING',(0,0),(-1,-1),8),('BOTTOMPADDING',(0,0),(-1,-1),8)]));story.append(t)
p('Adom Custom Electronic Load','Title');p('Engineering datasheet | R0 / Board34 + R10 prototype cooling | 13 October 2026','Heading2')
p('PROTOTYPE — NOT RELEASED FOR FABRICATION. Values below are design limits or software policy unless explicitly stated. No physical load accuracy, thermal performance, safety certification or industrial qualification is claimed.')
p('Digitally controlled constant-current sink for testing DC modules in an Adom probing workcell. Q1 converts DUT power to heat. Auxiliary 5 V independently powers logic and a tachometer-equipped fan.','BodyText')
table([['Parameter','Design / policy'],['DUT operating voltage','4.5–24 V DC; not an absolute-maximum or transient rating'],['Load mode','Constant current; no qualified constant-resistance or constant-power mode'],['Allowed setpoint','0 to min(3 A, 5 W / DUT voltage). Examples: 5 V → 1 A; 12 V → 0.417 A; 24 V → 0.208 A. The 3 A ceiling is not attainable within this 5 W voltage envelope.'],['Power ceiling','5 W initial operating target; not a measured continuous dissipation rating'],['Cooling review conditions','5.5 W Q1 screening allowance, 40 °C ambient, 55 °C case upper-bound ceiling; no fan cooling credit'],['Accuracy / dynamics','Unspecified pending calibration and physical load-step tests'],['Hardware status','Unbuilt / unflashed; STM32 bring-up firmware intentionally output-locked OFF']])
story.append(PageBreak());p('Interfaces and protection','Title')
table([['Connection','Assignment / behavior'],['J1 / J2','DUT positive / ground'],['J3 / J4','5V_AUX / ground. Regulated external auxiliary supply; supply-current requirement remains unqualified.'],['J8 USB-C','USB CDC control; VBUS detection only, not fan/logic power. No galvanic isolation claimed.'],['J5 / J6','3.3 V UART TX / RX alternative; not 5 V tolerant control wiring'],['J7','Probe ground contact J7 at native PCB (116, 102) mm; full modeled motion review passes; physical contact qualification remains open'],['J9 fan','1 = 5V_AUX; 2 = GND; 3 = FAN_TACH. Confirm assembly polarity before energizing.']])
p('Control protocol','Heading2');p('Draft adom.load/1 newline-delimited JSON: identify, claim, status, set_current, arm, off, clear_fault and heartbeat. One controlling owner, boot identity and increasing request IDs. Request/applied current are separate. Firmware qualification cannot be enabled by a serial command.')
p('Telemetry and interlocks','Heading2');p('Telemetry includes DUT voltage/current/power, case-temperature estimate and qualified upper bound, fan RPM and validity, setpoint, applied current, faults and sample age. Missing/stale values remain unknown. Hardware comparator/latch/watchdog protection and software lease/sensor checks require physical verification. Fault clearing does not automatically re-arm.')
p('Fan','Heading2');p('Sunon MF40200V1-1000U-G99: 5 V, nominal 7000 RPM ±15%, two tach pulses/revolution; 3.3 V / 10 kΩ tach pull-up. The fan runs with auxiliary power; there is no independent PWM speed command. Software requires qualified 5000–10000 RPM, eight valid periods, and rejects tach history older than 15 ms. These are policy values, not measured shutdown timings. RPM does not establish airflow.')
story.append(PageBreak());p('Mechanical, fabrication and qualification','Title')
p('Revision identity','Heading2');p('Board34 is the adopted electrical design (SHA256 74f8ac20cba6b95db26c40f5369156efec80cb80e35388fe064f6891ff78ff11). The displayed R10 cooling assembly uses its 72 x 104 mm, nominal 1.6 mm PCB: four large corner pins on a 64 x 96 mm span, USB-C at the edge, and routed TP14/TP15/TP16. Native DRC is clean, 107 live footprint poses match, and all 276 connected pins match the schematic. Do not manufacture from the hero model.')
p('Manufacturing intent','Heading2');p('Two copper layers using 3rdPartyFab. Standard finished copper: 1 oz; 2 oz available on request and requires corresponding rule/analysis updates. Assembly retention, fan clearance, wire routing, probe approach and tether clearance remain validation items. R10 has an outboard tach fan, reinforced carrier with 6 mm fan plate, and a Wakefield 125319 heatsink cut to 76.2 mm, through nominal AlN insulation and a copper spacer. Overall assembly is approximately 151 x 120 x 69 mm. The public fin model is an original simplified render; machining, print tolerances and retention require prototype review. The exact Board34/R10 thermal screen passes using catalog passive resistance with no heatsink fan credit.')
p('Evidence and open acceptance','Heading2');p('Board34 passes native routing DRC, exact-input current screening and schematic parity. R10 mechanical screening finds no unexpected intersections against fitted component bounds or in the fastener/support check, and the carrier is one connected solid; this is not structural qualification. The previous front support interfered with probing and has been replaced. The exact supplier-profile R10 model passes the full 20-touch replay with zero head collisions and zero snags; the original public-render model also passes its separate input-bound replay. Wire-on-heatsink wear, exact fin geometry, installed cooling, manufacturing release, firmware integration and physical calibration remain open. Fresh 0.1 mm screening predicts nominal Q1 case 52.07 C and junction 54.38 C at 40 C ambient and 5.5 W; margin is narrow and catalog/contact assumptions require later physical qualification. Simulated fan motion is not hardware evidence.')
p('Design sources','Heading2');p('Project source: https://wiki.adom.inc/adom/adom-custom-electronic-load<br/>docs/firmware-contract.md — protocol, limits and qualification locks<br/>docs/fan-feedback.md — fan and tach implementation/provenance<br/>docs/completion-plan.md — current revision boundaries<br/>probing-flow/progress.json — probing status (consult newest evidence)<br/>docs/cost/README.md — partial pricing, exclusions and unknowns')
p('Document control','Heading2');p('Engineering datasheet revision R0-20261013-BOARD34-R10. This is a project-authored specification, not a component manufacturer datasheet. Design values may change following prototype qualification. Refresh this document whenever board identity, limits, connectors or acceptance status changes.')
def footer(c,doc):c.setFont('Helvetica',8);c.drawString(45,25,'ADOM electronic load | Engineering prototype | R0-20261013');c.drawRightString(550,25,str(doc.page))