Selected sensor software binding
NTCLE428E3103H301A replaces the old NTCLE100 nominal conversion in the output-locked bring-up adapter. The original standalone candidate remains historical. The30 table rows and primary-document provenance are in design/sensor-candidate.json. checks.json covers the compiled curve and all4096 ADC codes; host-tests.log contains sanitizer-backed host and protocol results.
Outputs remain unchanged on failed conversion. The ADC adapter invalidates diagnostics until reset after saturated or out-of-table input. No calibration or interpolation error bound is claimed. The55 °C policy setting is unchanged and is not an assurance of55 °C actual case temperature. Hardware, thermal lag and calibration gates remain open.